Web Mail Tuesday, 22nd May 2018 06:52pm UK
Product comparison.
Feature Specification:
IDmicro
 
Mass range
1-70
 
Maximum acceleration voltage
5KV
 
Resolution CNOS
75
 
Magnet effective radius
90 deg 14 cm
 
H/D detector
No
 
Dispersion
140mm
 
Simultaneous mass range
Not applicable
 
Max number of collectors
3
 
Differential pumping
No
 
Sensitivity molecules CO2 / 44 ion in DI
Not applicable
 
Sensitivity molecules CO2 / 44 ion in CF
1100
 
Sample Consumption
< 0.045nmol / sec gives M44 of 5e-9A
 
Abundance sensitivity DI
Not applicable
 
Ion Source Linearity
0.02 / nA ion current (mass 44)
 
H3+ contribution
< 10ppm / nA
 
Continuous Flow Specifications:
IDmicro
 
Reference Gas Precision 1σ; n=10; 5nA
δ13C (CO2)
0.08 per mil
 
δ18O (CO2)
0.08 per mil
 
δ15N (N2)
0.08 per mil
 
δ15D (H2)
Not applicable
 
Reference Gas Linearitiry / nA
δ13C (CO2)
0.02 per mil
 
δ18O (CO2)
0.02 per mil
 
δ15N (N2)
0.02 per mil
 
δ15D (H2)
 
 
Installation Requirements:
IDmicro
 
Compressed air
No
 
Water chiller
No
 
Location
Bench Top
 
Power
Single phase
 
Weight
45Kg excluding pump
 
Dimensions
47 x 67 x 25 cm
 
© Copyright 2009-2010 Compact Science Systems     Terms     Privacy     Site produced & maintained by honest.